Product Overview
Features
  • 0.05% full scale (FS) accuracy
  • Pressure ranges to 20,000 psi (1,400 bar)
  • Eleven selectable pressure units
  • Large, easy-to-read display with a five-digit resolution
  • % pressure indication and additional bar graph for quick visual reference
  • Temperature-compensated accuracy from 14°F to 122°F (-10°C to 50°C)
  • 0 to 5V analog output
  • Pressure switch test
  • Minimum/maximum, tare, and alarm functions
  • IDOS compatible and RS232 serial interface
  • Free of charge monitoring and control software
  • Stainless steel or Inconel pressure cavity for aggressive media
Benefits
Accuracy

0.05% FS including non-linearity, hysteresis, repeatability, and temperature effects from 14°F to 122°F (-10°C to 50°C)

Burst Pressure

Burst pressure is 2x working pressure (exception, 20,000 psi (1,400 bar); burst pressure 29,000 psi (2,000 bar).

Selectable Pressure Units

kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O and mmHg

Display

Pressure reading: 5 digits with 0.5 in (12.7 mm character height

Full-scale indicator: 2.5 digits with 0.25 in (6.35 mm) character height.

Pressure Bar Graph

In addition to the numeric pressure indication, the DPI 104 LCD display contains a circular twenty segment bar graph to provide the user with a quick visual estimation of pressure from 0 to 100% full-scale output. The bar graph increments represent 5% of the user-selectable range.

Minimum/Maximum Values

最小和最大压力值可说played on the DPI 104. This function can be enabled/disabled, and reset by the user.

Switch Test

104 DPI功能开关测试函数will capture and display the open and closed values from an external pressure switch. Maximum switch impedance 200 Ω.

Voltage Output

The DPI 104 can be programmed to provide a 0 to 5 V output signal that can be configured as directly proportional to the pressure shown in the display or set to a fixed value. The voltage output mode provides 0.1% FS accuracy from 50 mV to 5 V.

Alarm Output

The alarm output consists of an open drain field-effect transistor (maximum current 250 mA, maximum voltage 24 VDC)

Adjustable Mounting Positions

For added convenience, the DPI 104 housing will rotate 320° around the pressure fitting and the faceplate can be rotated in any orientation for optimum visibility.

Menu Lock

To guard against unauthorized menu access, the DPI 104 features a menu and tare lock function

Pressure Ranges

The DPI 104 can accommodate pressureranges up to 20,000 psi (1,400 bar)with 0.05% full scale (FS) accuracy.

Datasheet

Datasheets -DPI 104



Contact Us | Support

For more information on Druck products or to reach a member of your local Druck team please click >here

You can contact the Druck Support Team >here


Get in touch with an expert.

Your request has been submitted.
Thank you for your interest. A specialist will be in touch with you shortly.